CVE-2024-48966: Life2000 service tools for test and calibration do not support user authentication
Noroxi assessment
Priority: Plan (action score 40/100)
- The sources we track (CISA KEV, Exploit-DB, Metasploit, Nuclei, GitHub) show no public exploit code or exploitation in the wild.
- Probability of exploitation in the next 30 days: 0.69% (FIRST EPSS); higher than 53.1% of all records.
- Anyone who can reach it over the internet can trigger it. No account or password is required. No user action is required. No special conditions are required; it is repeatable.
- According to CISA the attack is automatable: targets can be found by mass scanning.
- According to the record, no fixed version is available yet.
Add it to your regular patch cycle. Update when a fix is released; follow the vendor's advisory.
Generated automatically from the record's data: CVSS (NVD and CNA), CISA KEV and SSVC, FIRST EPSS, exploit sources, fixed versions. How we calculate it
Primary sources: NVD · CVE.org · FIRST EPSS
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
- Published
- Nov 14, 2024
- Updated
- Jun 17, 2026
- EPSS
- 0.6% · 53th percentile
- CWE
- CWE-306
Report tools
Action score
40
Plan
Add it to your regular patch cycle.
- CVSS
- 40 / 40 · 10.0 / 10
- CISA KEV
- 0 / 30 · Not listed
- EPSS
- 0 / 30 · 0.6%
CISA SSVC decision
- Exploitation
- none
- Automatable
- yes
- Technical impact
- total
Vulnrichment: CISA's decision-tree inputs.
CNA vs NVD score
- NVD
- —
- CNA · Baxter
- 10.0
- NVD has not scored this yet; the score shown is the CNA’s.
The score given by the assigning authority versus NVD’s independent score. A gap means the severity is contested.
Get told when this record changes
- Not in KEV: no record of exploitation in the wild
- No public exploit code on record
- No fixed version yet
If any of these changes, we'll tell you in the daily email digest.
Running WordPress? Scan your site free with WP Lens
Noroxi analysis
What does it mean in your environment?
The assessment above is generated automatically from the record's data; it doesn't know your environment. Our team verifies on your systems whether this vulnerability is actually exploitable and which mitigation is enough.
We use this product, ask for helpAffected systems
Is your version affected?
Pick the product and type the version you run. It's compared with this record's version ranges in your browser; the version isn't sent anywhere.
| Vendor | Product | CPE |
|---|---|---|
| baxter | life2000 ventilator firmware | cpe:2.3:o:baxter:life2000_ventilator_firmware |
Versions reported by the vendor
Affected version ranges reported by the assigning authority (Baxter). Independent of NVD's CPE analysis and usually ahead of it.
Baxter Life2000 Ventilation System
- 06.08.00.00 and prioraffected
Same product
baxter: all recordsOther highest-scoring records for the same primary product.
- CVE-2024-48967Life2000 ventilator and Service PC lack sufficient audit logging capabilities40Plan
- CVE-2024-48974Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates37Monitor
- CVE-2024-48973Debug port on Life2000 Ventilator serial interface is enabled by default37Monitor
Published before and after this one in the same product
- ← EarlierCVE-2024-9832Nov 14, 2024 · No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator37Monitor
- ← EarlierCVE-2024-9834Nov 14, 2024 · Improper data protection on Life2000 ventilator serial interface37Monitor
- Later →CVE-2024-48970Nov 14, 2024 · Life2000 Ventilator microcontroller lacks memory protection37Monitor
Remediation
Which version to upgrade to
Fix versions compiled from the vendor, package registries and Microsoft. Verify the vendor's note before upgrading.
No fix version is recorded for this entry. Check the references for vendor advisories.
Exploit status
No known public exploit
No public exploit has been observed yet. That doesn't mean you're safe, only that the bar is a little higher.
Research context
For pentesters and researchers: attack profile, score disagreement, timeline, patch commits, credits, variant and chain candidates, bug bounty scope. All derived from existing data; no exploit code.
Timeline
From publication to today: proof of concept, Metasploit module, CISA KEV and fix record. Dates are as reported by the sources.
No dated events beyond publication.
No history yet (it accumulates with the daily EPSS job).
Patch and commit links
Commit, PR and diff links among the references. A starting point for patch-diffing and variant hunting; fixes, not exploits.
No commit or PR link among the references.
Credits
All researchersFinders, reporters and analysts named in the CNA record. Click a name for that researcher’s other records.
No credits in the CNA record.
Variant candidates
Same product, same weakness class, within 18 months. If the patch missed the root cause, the sibling bug is here.
No nightly-computed relations.
Chain candidates
An authentication bypass and a privilege-requiring bug in the same product, published close together: combined they may become an unauthenticated path.
—
Bug bounty scope
No known public program.
Source: bounty-targets-data (public HackerOne, Bugcrowd, Intigriti, YesWeHack listings).
Technical details
Attack conditions
- Anyone who can reach it over the internet can trigger it.
- No account or password is required.
- No user action is required.
- No special conditions are required; it is repeatable.
If successful
- Confidentiality
- high · data can be read
- Integrity
- high · data or configuration can be modified
- Availability
- high · the service can be disrupted
Impact can extend beyond the vulnerable component (scope changes).
- Attack vector
- Network
- Attack complexity
- Low
- Privileges required
- None
- User interaction
- None
- Scope
- Changed
- Confidentiality impact
- High
- Integrity impact
- High
- Availability impact
- High
Weakness class (CWE)
CWE-306 · Missing Authentication for Critical FunctionCVSS vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
nvd-secondary
Root cause
A function added for development or support ships to production without authentication. It often relies on the assumption that it is “only reachable from the internal network.”
The bug in code
The whole CWE-306 classA representative example of this vulnerability class. Not the vendor's source code; it shows the faulty pattern and its fix.
Vulnerable
// Added for the support team
app.post("/maintenance/reset", resetDevice);Fixed
if (config.maintenanceEnabled) {
app.post(
"/maintenance/reset",
requireSession, requireRole("admin"), resetDevice
);
}Attack context
MITRE CAPEC attack patterns and ATT&CK techniques for this weakness class (CWE). A starting point for detection rules and threat hunting.
Attack patterns (CAPEC)
ATT&CK techniques
—
Change log
No changes recorded on tracked fields yet. Score, KEV, exploit maturity and fix status changes appear here.
References
Vendor advisories and official records. Exploit/PoC links are deliberately left out.