CVE-2018-10596: Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints
Noroxi assessment
Priority: Monitor (action score 28/100)
- The sources we track (CISA KEV, Exploit-DB, Metasploit, Nuclei, GitHub) show no public exploit code or exploitation in the wild.
- Probability of exploitation in the next 30 days: 1.3% (FIRST EPSS); higher than 71.5% of all records.
- Anyone who can reach it over the internet can trigger it. A low-privileged account is enough. A user must open a link or visit a page. Special conditions such as timing or configuration are required.
- According to the record, no fixed version is available yet.
- Mentioned 1 time in Turkish national cyber security advisories (siberguvenlik.gov.tr).
Low priority for now. Update when a fix is released; follow the vendor's advisory.
Generated automatically from the record's data: CVSS (NVD and CNA), CISA KEV and SSVC, FIRST EPSS, exploit sources, fixed versions. How we calculate it
Primary sources: NVD · CVE.org · FIRST EPSS
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
- Published
- Jul 2, 2018
- Updated
- Jun 16, 2026
- EPSS
- 1.3% · 71th percentile
- CWE
- CWE-923
Report tools
Action score
28
Low priority for now.
- CVSS
- 28 / 40 · 7.1 / 10
- CISA KEV
- 0 / 30 · Not listed
- EPSS
- 0 / 30 · 1.3%
CNA vs NVD score
- NVD
- —
- CNA · icscert
- 7.1
- NVD has not scored this yet; the score shown is the CNA’s.
The score given by the assigning authority versus NVD’s independent score. A gap means the severity is contested.
Get told when this record changes
- Not in KEV: no record of exploitation in the wild
- No public exploit code on record
- No fixed version yet
If any of these changes, we'll tell you in the daily email digest.
Is your version affected? Check itNoroxi analysis
What does it mean in your environment?
The assessment above is generated automatically from the record's data; it doesn't know your environment. Our team verifies on your systems whether this vulnerability is actually exploitable and which mitigation is enough.
We use this product, ask for helpAffected systems
Is your version affected?
Pick the product and type the version you run. It's compared with this record's version ranges in your browser; the version isn't sent anywhere.
| Vendor | Product | CPE |
|---|---|---|
| medtronic | 2090 carelink programmer firmware | cpe:2.3:o:medtronic:2090_carelink_programmer_firmware |
| medtronic | 2090 carelink programmer | cpe:2.3:h:medtronic:2090_carelink_programmer |
Versions reported by the vendor
Affected version ranges reported by the assigning authority (icscert). Independent of NVD's CPE analysis and usually ahead of it.
Medtronic 2090 CareLink Programmer
- All versionsaffected
Medtronic 29901 Encore Programmer
- All versionsaffected
Affected versions
NVD version ranges (for catalog products). Add the product to your stack with a version and matching uses these.
- medtronic 2090 carelink programmer firmwareall versions
Same product
medtronic: all recordsOther highest-scoring records for the same primary product.
- CVE-2018-5448Medtronic 2090 Carelink Programmer Relative Path Traversal19Monitor
- CVE-2018-5446Medtronic 2090 Carelink Programmer Storing Passwords in a Recoverable Format19Monitor
Remediation
Which version to upgrade to
Fix versions compiled from the vendor, package registries and Microsoft. Verify the vendor's note before upgrading.
No fix version is recorded for this entry. Check the references for vendor advisories.
Exploit status
No known public exploit
No public exploit has been observed yet. That doesn't mean you're safe, only that the bar is a little higher.
Research context
For pentesters and researchers: attack profile, score disagreement, timeline, patch commits, credits, variant and chain candidates, bug bounty scope. All derived from existing data; no exploit code.
Timeline
From publication to today: proof of concept, Metasploit module, CISA KEV and fix record. Dates are as reported by the sources.
No dated events beyond publication.
Credits
All researchersFinders, reporters and analysts named in the CNA record. Click a name for that researcher’s other records.
- Billy Rios and Jonathan ButtsWhitescope LLC identified these vulnerabilities and reported them to CISA.finder
National notice (Türkiye Cybersecurity Directorate / USOM)
Official security notices citing this record; remediation advice is on the agency's page.
- TR-20-066 · Jan 30, 2020(Medtronic Carelink Zafiyeti)
Not on record for this CVE: EPSS history · patch commit or PR link · variant or chain candidate · public bounty program.
Technical details
Attack conditions
- Anyone who can reach it over the internet can trigger it.
- A low-privileged account is enough.
- A user must open a link or visit a page.
- Special conditions such as timing or configuration are required.
If successful
- Confidentiality
- high · data can be read
- Integrity
- high · data or configuration can be modified
- Availability
- high · the service can be disrupted
- Attack vector
- Network
- Attack complexity
- High
- Privileges required
- Low
- User interaction
- Required
- Scope
- Unchanged
- Confidentiality impact
- High
- Integrity impact
- High
- Availability impact
- High
Weakness class (CWE)
CWE-923 · Improper Restriction of Communication Channel to Intended EndpointsCVSS vector
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
nvd-secondary
Attack context
MITRE CAPEC attack patterns and ATT&CK techniques for this weakness class (CWE). A starting point for detection rules and threat hunting.
Attack patterns (CAPEC)
Change log (0)
No changes recorded on tracked fields yet. Score, KEV, exploit maturity and fix status changes appear here.
References
Show all references (2 more)Show less
Vendor advisories and official records. Exploit/PoC links are deliberately left out.