An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, a

An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack.

MediumCVSS 5.9 · v3.0—No exploit No fix recorded
Published
Feb 15, 2018
Updated
Jun 16, 2026
EPSS
0.7% · 50th percentile
Follow this CVE

Sign in to follow · You’ll be notified if a followed record enters KEV, gets an exploit or is updated.

Report tools

JSON

Action score

23

Monitor

Low priority for now.

CVSS
23 / 40 · 5.9 / 10
CISA KEV
0 / 30 · Not listed
EPSS
0 / 30 · 0.7%

Noroxi analysis

No Noroxi analysis for this record yet

We don't hand-write analysis for the hundreds of thousands of vulnerabilities in the database; that wouldn't be honest. For notable, high-impact vulnerabilities our team writes the mechanism, detection and remediation steps.

We use this product, ask for help

Affected systems

VendorProduct
smiths-medicalmedfusion 4000 wireless syringe infusion pump

Affected versions

NVD version ranges (for catalog products). Add the product to your stack with a version and matching uses these.

  • smiths-medical medfusion 4000 wireless syringe infusion pump1.1
  • smiths-medical medfusion 4000 wireless syringe infusion pump1.5
  • smiths-medical medfusion 4000 wireless syringe infusion pump1.6

Versions reported by the vendor

Affected version ranges reported by the assigning authority (icscert). Independent of NVD's CPE analysis and usually ahead of it.

  • n/a Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

    • Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pumpaffected

Other highest-scoring records for the same primary product.

  • CVE-2017-12718A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.
    36Monitor
  • CVE-2017-12720An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.
    33Monitor
  • CVE-2017-12724A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and
    32Monitor
  • CVE-2017-12726A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6
    29Monitor
  • CVE-2017-12725A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and
    22Monitor
  • CVE-2017-12722An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.
    22Monitor

Remediation

Which version to upgrade to

Fix versions compiled from the vendor, package registries and Microsoft. Verify the vendor's note before upgrading.

No fix version is recorded for this entry. Check the references for vendor advisories.

Exploit status

No known public exploit

No public exploit has been observed yet. That doesn't mean you're safe, only that the bar is a little higher.

Research context

For pentesters and researchers: attack profile, score disagreement, timeline, patch commits, credits, variant and chain candidates, bug bounty scope. All derived from existing data; no exploit code.

Timeline

From publication to today: proof of concept, Metasploit module, CISA KEV and fix record. Dates are as reported by the sources.

No dated events beyond publication.

EPSS, last 120 days

FIRST EPSS daily score; only changes of 0.01 or more are recorded (step chart).

Patch and commit links

Commit, PR and diff links among the references. A starting point for patch-diffing and variant hunting; fixes, not exploits.

No commit or PR link among the references.

Finders, reporters and analysts named in the CNA record. Click a name for that researcher’s other records.

No credits in the CNA record.

Variant candidates

Same product, same weakness class, within 18 months. If the patch missed the root cause, the sibling bug is here.

No nightly-computed relations.

Chain candidates

An authentication bypass and a privilege-requiring bug in the same product, published close together: combined they may become an unauthenticated path.

—

Bug bounty scope

No known public program.

Source: bounty-targets-data (public HackerOne, Bugcrowd, Intigriti, YesWeHack listings).

Technical details

Attack conditions

  • Anyone who can reach it over the internet can trigger it.
  • No account or password is required.
  • No user action is required.
  • Special conditions such as timing or configuration are required.

If successful

Confidentiality
high · data can be read
Integrity
none
Availability
none
Attack vector
Network
Attack complexity
High
Privileges required
None
User interaction
None
Scope
Unchanged
Confidentiality impact
High
Integrity impact
None
Availability impact
None

CVSS vector

CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N

nvd-primary

Root cause

A certificate validation error is silently swallowed instead of terminating the connection. Often, code added for a test environment makes its way into production.

A representative example of this vulnerability class. Not the vendor's source code; it shows the faulty pattern and its fix.

Vulnerable

kotlin
override fun checkServerTrusted(chain: Array<X509Certificate>, type: String) {  // Left empty for the test environment}

Fixed

kotlin
val client = OkHttpClient.Builder()  .certificatePinner(    CertificatePinner.Builder()      .add("api.sirius.example", "sha256/…")      .build()  ).build()

Attack context

MITRE CAPEC attack patterns and ATT&CK techniques for this weakness class (CWE). A starting point for detection rules and threat hunting.

Change log

No changes recorded on tracked fields yet. Score, KEV, exploit maturity and fix status changes appear here.

References

Vendor advisories and official records. Exploit/PoC links are deliberately left out.